Kofloc RK1950AP Large Capacity Flow Meter

General Information

Kofloc RK1950AP Large Capacity Flow Meter Model Amber tapered tube with secure structure with Float receiver functions also as a joint, for secure piping and The block tapered tube has an O-ring seal. The flow meter (area type flow meter) has a float inside a conical cylinder (tapered tube) with the upper end larger than the lower end. The float moves up and down according to the intensity of flow, and the reading is taken when the float balances indicate the flow rate. KOFLOC provides flow meters with a valve that can perform both measurement and control of the flow rate, flow meters with an alarm function, and many other products.

The Principles of  Flow Meter Float Type

As shown in the figure, the area (float type) flow meter has a conical cylinder (tapered tube) with the upper end larger than the lower end, having a float inside. The float moves up and down according to the flow, and the reading taken when it balances indicates the flow rate.

Fig 1 Principle of Rotameter (Float Type)
Fig 1 Principle of Rotameter (Float Type)
  1. The symbols and units are determined as follows:
  • Vf: Volume of float [cm3]
  • ρf: Density of float [g/cm3]
  • af: Maximum cross-section of float [cm2]
  • P1: Pressure directly under float [Pa]
  • P2: Pressure directly above float [Pa]
  • ρ: Flow density [g/cm3]
  • U1: Flow velocity directly under float [cm/sec]
  • U2: Flow velocity in float clearance [cm/sec]
  • A1: Cross section directly under float [cm2]
  • A2: Cross section of float clearance [cm2]
  • Q: Flow rate [cm3/sec]
  • g: Gravitational acceleration
  1. The force that pushes up the float is af (P1–P2). Subtracting the buoyancy, the gravitation of the float is Vf (ρf–ρ) g.

Flow rate when load pressure is applied (Measurement of gas)1.

As shown in Figure A, the float-type flow meter is generally used with the rear section of the fl ow meter exposed to the atmosphere or without any load (without pressure loss resistance). The scale of such a flow meter is called the “atmospheric scale.”

Fig A - Principle of Rotameter (Float Type)
Fig A – Principle of Rotameter (Float Type)

Flow rate when load pressure is applied (Measurement of gas) 2.

The scale of such a flow meter is called the “atmospheric scale.” In fact, however, there often exists load pressure resistance as shown in Figure B. As shown in the example of the principle, load pressure resistance does not allow the actual flow rate to be indicated. There is a method to find a rough standard using the equation for calculation, but some errors will be caused because of the gap between theory and practice. Such an error will be caused both in the pressurized condition and in the vacuum condition.

Fig B - Principle of Rotameter (Float Type)
Fig B – Principle of Rotameter (Float Type)

For example, the flow control valve and the like are arranged in the later stage as shown in Figure C to control the pressure applied to the flow meter with a pressure regulator or the like. The scale of the flow meter is calibrated under that pressure condition. The scale of such a flow meter is called a “load pressure scale.” An example of using the load pressure scale is our GM Series gas mixing equipment. In the GM Series, the basic flow sheet is used as shown in Figure C so that the gas obtained with a constant flow rate will be pressurized.

The load pressure scale of a flow meter with a valve on the outlet side (needle, etc. on top of RK-1250) is automatically calibrated at all times, but if the flow meter will be used independently, please inform us of this in advance and select the Panotation. An extra charge will be added for load pressure calibration.

Fig C - Principle of Rotameter (Float Type)
Fig C – Principle of Rotameter (Float Type)

The upper section of a fl ow meter is on the outlet side, while the lower section is on the inlet side.

The equivalent flow chart of the upper and lower valves is shown in the figure.

The pressure loss of a flow meter is very small, not causing any problems when a flow meter is used individually. In the case of a flow meter with a valve, however, the inlet side pressure and outlet side pressure are important. The valve installed in a flow meter causes a pressure loss, generating a difference between the inlet pressure and outlet pressure. As shown on page 50, the reading of a float-type flow meter differs according to the applied pressure.

Therefore, for flow meters with a valve, care is needed to see what will be the applied pressure. In such cases, the inlet pressure P1 is throttled by a needle valve, changing into P2 at the outlet. n the lower section, this P2 is the pressure applied to the flow meter. When there is no significant load resistance (similar to atmospheric conditions) at the outlet and the subsequent section, P2 can be regarded as 0 MPa · G (gauge pressure), and as no pressure is applied to the flow meter, a general flow meter is used for calibration. (Refer to the atmospheric scale on page 50.) Since operating pressure conditions are necessary for valves, however, it is necessary to clearly indicate P1 and P2 when placing an order. Please indicate P2 as well when there is load resistance (when P2 pressure exists).

Fig D - Principle of Rotameter (Float Type)
Fig D – Principle of Rotameter (Float Type)

In the case of the upper valve, the inlet pressure P1 is applied to the flow meter.

In that case, P1 and P2 are necessary when selecting a valve as mentioned above, and the scale of the flow meter should be the negative pressure scale (P.50) of P1. Therefore, the pressure is controlled by a pressure regulator so that P1 will not change while us-ing a flow meter in general. Especially, when pressure is reduced to a vacuum in the rear stage, the flow reading will be incorrect unless an upper valve type, which reduces pressure P1 so that the flow meter will not be decompressed, is used. Care is needed.

Kofloc RK1950AP Large Capacity Flow Meter

Kofloc RK1950AP Large Capacity Flow Meter Model Amber tapered tube with secure structure with Float receiver functions also as a joint, for secure piping and The block tapered tube has an O-ring seal.

Features

  • Float receiver functions also as a joint, for secure piping.
  • The block tapered tube has an O-ring seal.
  • The integral housing protects the tapered tube.
  • The tapered tube made of engineering plastic is impact-resistant.

Standard  Specifications

  • Accuracy: ±2%F.S
  • Max. operating temperature: 120℃
  • Max. operating pressure: 1 MPa for 40 A or less 0.8 MPa for 50 A or more
  • The direction of flow: Bottom → Top
  • Coating :
    • *Finish Color: Gray (Close to Munsell N-6.5)
    • *Epoxy powder baking
  • Connection standard:
    • *JIS10K-RF 10A-100A RC female thread 3/8–1 inch
    • *Flange only ANSI150#RF, DIN
  • Part name Material
    • Float receiver: SUS304
    • O-ring: NBR
    • Tapered tube: Polysulfone resin (Transparent amber)
    • Housing: Ductile cast iron: FCD400
    • Rod: SUS304
    • Float: SUS304

Source : www.kofloc.co.jp