Categories: Gas AnalyzerSensor

PCA® 400 Combustion & Emissions Analyzer Bacharach

Bacharach’s PCA® 400 is a handheld versatile combustion and emissions analyzer for commercial and industrial applications. The analyzer may be configured for up to 4 gas sensors, including O2, CO, NO, NO2 and SO2. The

Bacharach PCA® 400’s portable and durable design is ideal for testing, tuning, and maintaining burners, engines, furnaces, and a variety of other commercial and industrial combustion applications.

The analyzer also features Bluetooth® communications that allow users to view real-time data send customizable reports from their compatible smartphone.

FEATURES

BENEFITS

Up to 4 Gas Sensors (O2, CO, NO, NO2 & SO2) Versatile analyzer for many commercial and industrial applications
Sample Conditioner Superior accuracy for low NOx and SOx applications
Real-Time PC Software View and save live data for easy compliance reporting
Combustion App Quickly create custom reports with comments and send via email (available for Android & iOS).
Long Life O2 Sensor Reduced maintenance cost and downtime
Automatic Dilution and Sensor Protection Protects sensors in the toughest combustion environments, extending life
Dual Bluetooth and IR Printer with magnet Print combustion and emissions reports in the field
Precalibrated, B-Smart® Sensors Quick and easy sensor replacement in the field & reduced downtime (doesn’t require cal. gas)

Specification

Specification Description
O2 0 to 20.9%
CO (high range) 0 to 40,000 ppm
NO2 0 to 500 ppm
CO-H2 0 to 10,000 ppm
NO 0 to 3,000 ppm
SO2 0 to 5,000 ppm
Fuels[1] Natural Gas, Coal, Oil 2/4/6, Propane, Wood / Biofuels,
Kerosene, Bagasse, Digester Gas, B5, Pellets, KOKS, LEG,
LPG, Butane and Wood Chips
Environmental
Conditions
Stack Temperature: -4 to 2192ºF (-20 to 1200 ºC)
Ambient Temperature: -4 to 999 ºF (-20 to 537 ºC)
Draft / Differential Pressure: ± 72 inch of H2O (± 179mB)
Display 4.3“ (10.9 cm) backlit, color LCD with resistive touch interface
Display Languages English, Spanish, French
Power Rechargeable Li-Ion Battery ≈ 12 hours
4 × AA Disposable Lithium Batteries ≈ 12 hours
4 × AA Disposable Alkaline Batteries ≈ 5 hours
AC Adapter (100-240VAC @ 47-63 Hz) ≈ continuous
Size (H × W × D) 10″ × 3.8″ × 2.5″ (25.4 × 9.7 × 6.4 cm)
Weight 1.5 lbs. (0.68 kg) w/ Li-Ion battery pack
Operating Conditions 32 to 113°F (0 to 45°C), 15 to 90% RH (non-condensing), 1 ATM ± 10%
Warm-up Time 60 seconds
Memory 500 records
Communications Bluetooth® 4.0, USB 2.0 (micro-B), IrDA
Approvals CE & EN50379 Parts 1 and 2 (Siegert only)
Sample Flow Rate 0.6 to 0.7 L/min (standard)
1.0 L/min (high flow mode)

Model

Part Number Description
2411-1112 PCA® 400 (analyzer with O2, CO and NO sensors, 12″ probe, 7.5′ sample hose assembly, IrDA + Bluetooth® wireless printer and rugged protective carrying case)
2412-1312 PCA® 400 for NOx Measurement (analyzer with O2, CO, NO and NO2 sensors, 12″ probe, 7.5’ Viton™ sample hose assembly, IrDA + Bluetooth® wireless printer and rugged protective carrying case))
2413-1312 PCA® 400 for SOx Measurement (analyzer with O2, CO, NO and SO2 sensors, 12″ probe, 7.5’ Viton™ sample hose assembly, IrDA + Bluetooth® wireless printer and rugged protective carrying case)

 

Rudy Wiratama