Hengesbach Pressure and Level Transmitter Keramess KS 100/101

Pressure and Level Transmitter Keramess KS 100/101
Pressure and Level Transmitter Keramess KS 100/101

The KERAMESS pressure transmitters are suitable for taking pressure and filling level measurements in pipelines and containers. The wide range of process connections enables use in all applications in the food and pharmaceutical industries. The vacuum-proof and extremely overload safe measuring cell with a ceramic membrane made of highly pure aluminium oxide works on the basis of the capacitive measuring principle. The KERAMESS pressure transmitters are designed to measure from -1/0…1 to -1/0…70bar. The measuring ranges 0…0.05bar and -0.1…+0.1bar are available alongside other special measuring ranges.

Thanks to the ceramic membrane, the KERAMESS pressure transmitters are particularly suited for aggressive and abrasive media. The design for long-term medium temperatures of up to 125°C enables CIP and SIP cleaning methods to be used on the transmitters. The high protection classes of IP67 and IP69K guarantee that the devices can be safely cleaned on the outside with foam and a high-pressure cleaner and that moisture is reliably prevented from entering into the device. For additional protection against moisture, the electronics are fully encapsulated in the housing. All the pressure transmitters in series 100/101 are designed for universal applications and, in particular, for applications with constantly high temperatures of up to 125°C. Furthermore, using the on-site display with series 100 and the display and operating module OPUSM in series 101, the pressure transmitters can be simply read out, configured and diagnosed.

Features:

  • precise measurements with an accuracy of ≤ ± 0,15% FS
  • with dry ceramic measuring cell
  • Turn Down 10
  • easy to clean an high protection classes IP 67 und IP69K
  • integrated on-site display or external OPUSM display and operating module for parametrization and display of measuring values
  • application strengths: measurement of content and process pressure with small measuring ranges with an overload-safe ceramic measuring cell, also suitable for abrasive media
  • direct entry of parameters with plain text display

Another Article:

Specification:

 

Download

General details

Device type / measuring principle KS 100/101: capacitive

Input

Measuring ranges KS 100/101
Standard nominal measuring ranges [bar] relative OP relative OP absolute OP
OP = overload protection [bar] 0.05 4 40 60 0.1 4
0.1 4 70 105 0.2 6
±0.1 4 -1…1 10 0.4 6
0.2 6 -1…2 18 1 10
0.4 6 -1…4 25 2 18
Special measuring ranges are available on request. 1 10 -1…10 40 4 25
2 18 -1…20 40 10 40
4 25 -1…40 60 20 40
All measuring cells are vacuum-proof
10 40 -1…70 105 40 60
20 40 70 105
Setting the measuring ranges Via the keypad of the OPUSM display and operating module / via the integrated on-site display
Setting ranges Start the measuring zero: 0…90% of the sensor’s nominal measuring span TD=10
Measuring span span: 10…100%  of the sensor’s nominal measuring span
Burst pressure DIN16086 ≥ 4-fold measuring range

Output

Output signal 2-wire:   4…20mA with a test circuit connection in the device
Fault signal Optional: 3.8mA, 22mA, hold (i.e. holding the last value)
Current limitation 3.85mA and 21.5mA (normal operation)
Integration time Continuously selectable between 0 and 300s (setting time after a pressure leap)

Measuring accuracy

Reference conditions acc. to DIN IEC 770
Linearity, hysteresis and repeatability as per the limit point method DIN IEC 770 ≤ ± 0.15% of the sensor’s nominal measuring range
Activation time < 5s (the device will carry out a self-test.)
Setting time (without damping) < 200ms
Long-time drift ≤ 0.2% of the span per year
Thermal hysteresis ≤ ± 0.75% beginning of the measuring range / ≤ ± 0.8% end of the measuring range

Conditions of use

Installation position / calibration position Any position / standing vertically (position-dependent zero point displacement)
Medium temperature T1: -40…+125°C (140°C over one hour at the most)
Ambient storage temperature Type 101: -40…+85°C
Type 100: -30…+75°C
(Below -20°C cable breakage might occur and the display’s function may be impaired.)
Protection class acc. to EN60529 IP 67 and IP 69K
Electromagnetic compatibility Sensitivity against interference:  acc. to DIN IEC 61000-6-2
Interference radiation: acc. to DIN IEC 61000-6-4

Construction

Electrical connection – Standard: cable screw connection M16x1.5, nickel-plated brass, stainless steel available on request
– Optional: round plug-in connector M12x1, nickel-plated brass, stainless steel available on request
– Optional: angle plug acc. to EN 175301-803
– Optional: reference cable
Process connection – All standard front-mounted process connections and those that are commonly used by the manufacturer
Materials – Field housing / lid: CrNiSt 1.4301 (304)
– Housing seal: FPM (Viton®)
– Pressure compensation element: polyamide
– Inspection gauge (type 100): polycarbonate
– Process connection: CrNiSt 1.4404 (316L)
– Process membrane: Al2O3 (99%)
– Locking screw (type 101): CrNiSt 1.4301 (304)
– Reference cable: 5-wire with reference tube: PUR (recommended: 80m maximum)

Display and operation

Display LCD, 4-digit numerical display and 5-digit alphanumerical display
Type 100: integrated on-site display (cannot be separated from device)
Type 101: external OPUSM display and operating module
Displayable units Pressure: mbar, bar, psi, Pa, mH2O, mmHg, Torr, atm, at, kg/cm²
Volume: l, hl, dm³, m³, ft³, US gal, UK gal, US bl, UK bl
Mass: kg, t, lbs, tn. sh., tn. l.
Additional displays Output current in mA or % (in relation to the span)
Operation Type 100: via the configuration menu with the integrated on-site display
Type 101:via the configuration menu with the external OPUSM display and operating module

Auxiliary energy resources

Power supply / burden 12-36V DC, max. burden: (Vsupply – 12V) / 24mA

Accessories 100 series

OPUSM display and operating module External display and operating module, CrNiSt, IP 67, 41x70mm, 1m connection cable and round plug-in
connector M12x1
Certificates Calibration certificate
Declaration of conformity
Material inspection certificates as per EN 10204

Data Sheet: Pressure and Level Transmitter Keramess KS 100/101

Ref: Hengesbach.com