Signet 2580 FlowtraMag® Meter

Signet 2580 FlowtraMag® Meter
2580 FlowtraMag® Meter

The GF Signet 2580 FlowtraMag is a full-bore plastic PVC in line style magnetic flowmeter. The PVC body with titanium or Hastelloy® C electrodes has no moving parts, and is two to three times lighter in weight compared to traditional metal magmeters on the market. It is designed for high accuracy flow measurement in short pipe runs, making it an ideal solution for industrial applications where performance and ease of use are important.

The FlowtraMag design is factory calibrated with certificate at ±1% of reading accuracy. It is offered in corrosion resistant materials to provide long-term reliability with minimal maintenance costs. The LED indicators show at-a-glance system status, including normal operation, zero flow and partially filled pipe detection.

The flow meter provides three different outputs; field selectable frequency or digital (S3L) as well as analog 4 to 20 mA. The FlowtraMag is capable of temperature readings of the media using the Signet 0252 Configuration Tool or GF Config Tool Bluetooth® app. The GF Config Tool Bluetooth® app supports iOS and Android for simple on-the-fly user configuration.

These versatile, easy-to-install meters deliver accurate flow measurement in pipe sizes of DN25 (1 in.), DN50 (2 in.) and DN100 (4 in.), optimized for performance in short pipe runs often associated with final effluent lines, well heads and water treatment skids.

     Another Article:

Features

Features

  • No moving parts
  • Lighter in weight compared to traditional metal magmeters
  • Reduced straight run requirements, ideal for final effluent lines, wellheads and skids
  • Factory calibrated with certificate (±1% of reading accuracy)
  • Partially filled pipe detection status indicator
  • Visual LED indicators make sensor status clear and easy to read
  • Reverse flow direction configurable with 0252 Configuration Tool or GF Config Tool Bluetooth® App
  • Capable of temperature readings of the media using the 0252 Configuration Tool or GF Config Tool Bluetooth® App
  • One device with three different outputs: field selectable Frequency or Digital (S³L), and analog 4 to 20 mA
  • On-the-fly configuration with GF Config Tool Bluetooth® App
  • Bluetooth® 4.2 capable, support iOS and Android for simple user configuration

Applications

  • Chemical Processing/Production
  • Cooling Tower
  • Filtration Systems
  • Water and Wastewater Treatment
  • Municipal and Industrial Water Distribution
  • Pool and Aquatics
  • Process Control, Water Process Flow
  • Reverse Osmosis
  • Scrubber Systems
  • Metal Recovery and Landfill Leachate
  • Mining

Specifications

General
Pipe Size Range DN25, DN50, DN100 in., 2 in., 4 in.
Flow Range Titanium or Hastelloy C  0.02 to 10 m/s 0.07 to 33 ft/s
DN25 (1″) 0.53 LPM to 266.35 LPM 0.14 to 70.36 GPM
DN50 (2″) 2.23 LPM to 1112.60 LPM 0.59 to 293.92 GPM
DN100 (4″) 8.72 LPM to 4357.83 LPM 2.30 to 1151.22 GPM
Minimum Conductivity μS/cm – water based
Power Cable Wire 7.6 m (25 ft) 2-conductor shielded
Output Cable Wire  7.6 m (25 ft) 5-conductor shielded
Cable wires may be extended up to 300 m (1000 ft), special order only
Wetted Materials
Flow Tube Body PVC
Electrode Titanium, grade 2 or Hastelloy C-276
O-rings FKM
Performance
Accuracy ± 1% of reading plus ± 0.01 m/s (0.033 ft/s), reference condition 50 μS/cm and water based
Repeatability ± 0.5% of reading @ 25 °C (77 °F)
Low Flow Cutoff 0.02 m/s (0.07 ft/s) (adjustable via 0252 Confi guration Tool or GF Confi g Tool App)
Electrical
DC Power (Functional Rating) 24 VDC, max 24 W (12 to 32 VDC)
Reverse Polarity Protected Up to 35 VDC
Over-Voltage Maximum Rating 35 VDC
Current Output
Passive (low power) 4 to 20 mA per ANSI-ISA 50.00.01 Class H
Active Output 4 to 20 mA
Passive Loop Voltage 12 to 32 VDC
Loop Accuracy ±32 μA (@ 25 °C @ 24 VDC)
Loop Resolution 5 μA
Loop Span 3.8 mA to 21 mA
Error Condition None, 3.6 mA or 22 mA
Max. Cable 300 m (1000 ft)
Max. Loop Resistance 600 Ω @ 24 VDC
Compatible with PLC, PC or similar equipment
Frequency Output
Frequency 5 to 24 VDC, 50 mA max.
Frequency Range 0 to 1500 Hz
Max. Cable 300 m (1000 ft)
Max. Pull-up Voltage 30 VDC, 10k pull-up recommended
Compatible with Signet 8900, 9900, 9950, and 0486 Profi bus Concentrator
Digital (S3L) Output
Digital (S3L) 4.5 to 5.5 VDC
Serial ASCII, TTL level 9600 bps Compatible with Signet 8900, 9900, 9950 and 0486 Profi bus Concentrator
Max. Cable Length Application dependent
Environmental Requirements
Enclosure NEMA 4X / IP65
Relative Humidity 0 to 95% non-condensing
Storage Temperature -10 °C to 60 °C (14 °F to 140 °F)
Operating Temperature
Ambient -10 °C to 60 °C 14 °F to 140 °F
Media 0 °C to 60 °C 32 °F to 140 °F
UL Environmental Rating UL 50, Type 6P Storage
Altitude 4,000 m 13,123 ft
Pressure/Temperature Ratings
Maximum Operating Pressure 10 bar @ 23 °C (145 psi @ 73 °F)
DN25 (1″) and DN50 (2″) 3.5 bar @ 60 °C (51 psi @ 140 °F)
DN100 (4″) 2.27 bar @ 60 °C (33 psi @ 140 °F)
Shipping Weight -Titanium or Hastelloy C
DN25 (1″) 3.4 kg 7.5 lbs
DN50 (2″) 4.5 kg 9.9 lbs
DN100 (4″) 8.3 kg 18.3 lbs
Standards and Approvals
CE, FCC, NSF (Titanium only, does not include fl ange gaskets)
UL, CUL  Recognized Component
RoHS compliant, China RoHS
Manufactured under ISO 9001 for Quality, ISO 14001 for Environmental Management and OHSAS 18001 for Occupational Health and Safety.

Download Datasheet : Signet 2580 FlowtraMag® Meter

Reference : gfps.com