Kimo CP210-R – Differential pressure and temperature sensor

About Transmitter

The CP 210-R measures the differential pressure of air (and neutral gases), and is ideal for use in industrial and pharmaceutical applications.

  • This unit is available in
    • Measuring range:
      • CP 211: -100 to +100 Pa
      • CP 212: -1,000 to +1,000 Pa
      • CP 213: -10,000 to +10,000 Pa
    • Power supply 24 Vdc/Vac or 100-240 Vac
    • Solenoid valve for auto-calibration (CP211/212)
    • With or woithout display

Features

  • 0-5/10 V or 0/4-20 mA outputs
    Two relays
    Quick and easy mounting with the 1/4 turn system with wall-mount plate
    Deux sorties relais
    Automatic autozero
    Trend indicator
    “¼ turn” system mounting with wall-mount plate
    Configurable intermediate ranges
    Input Pt100 on terminal block for temperature measurement, range from -100 to +400°C (probes as option)
Transmitter CP 210-R - Differential pressure and temperature sensor
Transmitter CP 210-R – Differential pressure and temperature sensor

Applications  Of  CP 210-R – Differential pressure and temperature sensor

  • Pressure measurement, etc. of various types of liquids and gasses.
  • The absolute pressure transmitter is widely used wherever elevation above altimeter measurement is required, such as navigation systems.
  • Nowadays, these transmitters are also contributing to the weather forecasting field for barometric measurements. They allow weather stations to be miniaturized. For instance, one can forecast the weather using tiny sensors in smartphones or tablets.
  • Most of the industrial applications demand a partial vacuum. This applies in case of food packaging systems, where residual pressure determines the life of the inside product. The absolute pressure sensor helps ensure the same pressure in each package.
  • Absolute pressure transmitters are also used in the automation industry. In petrol and diesel vehicles,

Benefits  &  Features  of  CP 210-R – Differential pressure and temperature sensor

  • Measuring range: CP 211: -100 to +100 Pa, CP 212: -1,000 to +1,000 Pa, and CP 213: -10,000 to +10,000 Pa
  •  Solenoid valve for auto-calibration (CP211/212)
  •  0-5/10 V or 0/4-20 mA outputs
  •  Two relays
  •  Automatic autozero
  •  Quick and easy mounting with the 1/4 turn system with wall-mount plate

Technical Information

General Features
Power supply24 Vac / Vdc ±10%
100-240 Vac, 50-60 Hz
Warning: risk of electric shock
Output2 x 4-20 mA or 2 x 0-20 mA or 2 x 0-5 V or 2 x 0-10 V (4 wires)
Common mode voltage <30 vac=”” br=””>Maximum load: 500 Ohms (0/4-20 mA)
Minimum load: 1 K Ohms (0-5/10 V)
Relay outputs2 changeover relays
NO: 5A / NC: 3A / 240 Vac
Galvanic IsolationInputs and outputs (100-240 Vac models)
Device fully protected by
DOUBLE ISOLATION or REINFORCED ISOLATION
Outputs (24 Vac/Vdc models)
ConsumptionCP210-B: 6 VA / CP210-H: 8 VA
Electrical ConnectionScrew terminal block for cables 2.5 mm2. Carried out according
to the code of good practice
PC communicationUSB-Mini Din cable
EnvironmentAir and neutral gases
AutozeroManual by push-button; Automatic by solenoid valve
(only CP211/CP212)
Type of fluidAir and neutral gases
Conditions of use (°C/%RH/m)From -10 to +50°C. In non-condensing condition.
From 0 to 2000 m.
Storage TemperatureFrom -10 to +70°C
SecurityProtection class II; Pollution degree 2;
Overvoltage category 2 (OVCII)
European directives2014/30/EU EMC; 2014/35/EU Low Voltage;
2011/65/EU RoHS II; 2012/19/EU WEEE
Housing Features
MaterialABS V0 according to UL94
ProtectionIP65
Display75 x 40 mm, LCD 19 digits 2 lines
Height of digits:
Values: 10 mm; Units: 5 mm
ConnectionRibbed Ø6.2 mm (CP211/212/213)
Pass-throughFor cables Ø6 mm maximum
Cable glandFor cables Ø8 mm maximum
Weight340g

 

Configurable intermediate or centre zero ranges

TransmittersPressure RangeAir Velocity Range*
CP211-100/+100 PaFrom 3 to 10 m/s
CP212-1000/+1000 PaFrom 3 to 30 m/s
CP213-10 000/+10 000 PaFrom 3 to 85 m/s

Source : kimoinstruments.com

Read More Article :

 