Bamo Turbidity Underwater Settled

Description

Bamo Turbidity Underwater Settled includes a MUDSonic TS immersion probe and a MUDSONIC MF operating unit. The detection is done between the emitter and receiver of ultrasonic waves. The probe evaluates the concentration of solid particles by attenuation of the signal. The probe is immersed at a fixed height in the tank; It triggers a switch when the concentration (predefined) is reached (sludge level rising by settlement). The trigger point and timer are adjustable on-site (real operating conditions).

Bamo Turbidity Underwater Settled Features

  • Detection of settled solid particles in water
  • Level monitoring with ultrasonic detector
  • Output: 2 limit contacts, 1 fault contact
  • Adjustable timer
  • Adjustment along 3 measuring ranges (For sludge densities within 50 to 250 g/l )

Application

  • Iron filings detection
  • Salt detection in the brine tank
  • Sludge height monitoring in Thickeners Primary clarifiers Industrial processes (sludges with iron hydroxide, high-density slurry, etc.)

Bamo Turbidity Underwater Settled Specification

Operating unit MUDSonic MF
Power supply100 to 255 V AC – 50/60 Hz (model MF G) or
10 to 30 V DC and 12 to 24 V AC (model MF D)
Consumption1…5W
Ambient temperature-10 … +45 °C
Relay outputs2 contacts, 1 threshold, potential-free (N.O.)
(Limit exceeded, below the limit)
1 potential-free contact, fault alarm (N.O.)
When the supply voltage is off or in case of power failure, all contacts are open.
Switching power250 V AC, 3A / 30 V DC, 1 A
Caution: The contacts are not protected against overload.
Display:2½ digit LED, 5×7 dot matrix display
1 Blue LED, relay energized
3 Yellow LEDs, measuring range (LOW / Medium / HIGH)
Settings:Through a rotary switch on the front
Delay:Adjustable from 0.1 up to 9.9 s
Probe MUDSonic TS
Detector material:PVC
Cable:TPK, 10 m long, shielded, 2 x1 mm²
Other cable materials and lengths requested ( PVC, oil, and hydrocarbons-proof, etc.).
(Max. length 50 m)
Dimensions145 x 90 x 25 mm
Protection:IP 68 acc. EN 60529
Measuring principle:Ultrasonic signal attenuation
Operating temperature-20 … +45 °C

Download Data Sheet: Turbidity Underwater Settled Solids Detector MUDSonic

Ref: bammo.eu