Aquaread AP-Lite Quality Water

General Information

Aquaread AP-Lite Quality Water is a simple probe with a single optical socket. This socket is able to house any of our optical electrodes, including turbidity and chlorophyll. A temperature sensor is also included on the probe. The AP-LITE package includes our rugged 3m cable, our GPS Aquameter, a range of accessories and a rugged carry case.

Whether you require a turbidity meter or handheld device for the measurement of Blue Green Algae in reservoirs, the AP-LITE is the portable water quality meter for you. Sometimes multi-parameter measurement is not required, however the flexibility to change sensors is. Our AP-LITE solves this with its single parameter, removable sensor design. Simply remove your current sensor, install the new sensor and calibrate, you are then ready to take measurements.

Aquaread AP-Lite Quality Water Features

  • IP67 handheld display
  • Backlit display with ergonomic keypad
  • 1110-point data memory
  • GPS recording and site tag information
  • Data export via USB to AquaLink utility
  • User replaceable AA batteries
  • Easy to follow calibration procedures

Aquaread AP-Lite Quality Water Package

As a package, our AP-LITE comes supplied with everything you will need to start taking measurements. This includes our GPS Aquameter portable meter, offered with its rugged IP67 design, visual display, and large onboard data logger for taking sports recordings, or data logging purposes. On-board GPS also allows you to record where measurements have been recorded and the water quality data.

Aquaread AP-Lite Quality Water Sensors

Our AP-LITE digital water quality sensor utilises the latest technology and allows you to connect any of our Aquaprobe optical water quality sensors. These sensors are designed and manufactured by Aquaread, to provide the best possible accuracy and performance, while remaining cost effective. The optical sensors can be easily removed by hand, making them easy to change when been used in the field. The AP-LITE offers sensor for the measurement of :

  • Turbidity,
  • Chlorophyll,
  • Blue-Green Algae,
  • Rhodamine,
  • Fluorescein,
  • Refined Oil,
  • CDOM / FDOM

Aquaread AP-Lite Quality Water Specifications

  • Protection Class : IP68 (permanent immersion)
  • Immersion Depth : Min 75mm. Max 100m**
  • Operating Temperature : -5?C – +70?C
  • Dimensions (L x Dia) : 250mm x 24mm
  • Weight : 400g

AP-LITE Optical Sensor Specifications

OpticalRange0 – 3000 NTU
Resolution2 Auto-range scales: 0.0 – 99.9 NTU, 100 – 3000 NTU
TurbidityAccuracy± 5% of auto-ranged scale
Range0 – 500.0 µg/L (ppb)
Resolution2 Auto-range scales: 0.00 – 99.99 µg/L, 100.0 – 500.0 µg/L
ChlorophyllRepeatability± 5% of reading
PhycocyaninRange0 – 300,000 cells/mL
Resolution1 cell/mL
(freshwater BGA)Repeatability± 10% of reading
PhycerythrinRange200,000 cells/mL
Resolution1 cell/mL
(marine BGA)Repeatability± 10% of reading
Rhodamine WT DyeRange0 – 500 µg/L (ppb)
Resolution2 Auto-range scales: 0.00 – 99.99 µg/L, 100.0 – 500.0 µg/L
Accuracy± 5% of reading
Fluorescein DyeRange0 – 500 µg/L (ppb)
Resolution2 Auto-range scales: 0.00 – 99.99 µg/L, 100.0 – 500.0 µg/L
Accuracy± 5% of reading
Refined OilRange0 – 10,000 µg/L (ppb) (Napthalene)
Resolution0.1 µg/L
Repeatability± 10% of reading
Range0 – 20,000 µg/L (ppb) (Quinine Sulphate)
Resolution2 Auto-range scales: 0.0 – 9,999.9 µg/L, 10,000 – 20,000 µg/L
CDOM / FDOMRepeatability± 10% of reading