Kofloc 8500 Series Mass Flow Meter

Introduction

Kofloc 8500 Series Mass Flow Meter/meter driven by 24VDC power supply has been developed as a successor of the MODEL 8300. The viewpoint change function of the display unit and the pattern setting function are unique to this model, and noise resistance has been improved dramatically. A sister model with a detachable display and setting unit is also available.

  • The high-lift actuator allows this compact model to control a large flow rate.
  • Equipped with a display and setting unit, this model can be operated by the 24VDC power supply.
  • The RS232C/RS485 communication function and integration function are provided as standard equipment.
  • The 14-bit converter permits to display and operation in 4-1/2 digits.
  • Control of the flow rate of inflammable gas is possible because the heat generating part of the sensor is not exposed to the gas.
  • There are no limitations to the mounting position that may be employed.
  • In addition to the SV setting, five other patterns can be set.
  • Auto zero and auto close functions are also standard.

Kofloc 8500 Series Specification

ITEM8500MC/MM8550MC/MM
Flow range10SCCM?20SLM30SLM 100SLM
N2 equivalent20/1ATMfreely selectablefreely selectable
Sensor typeThermal sensor
Valve type 1Normally closed proportional solenoid valve
Control range 12100 F.S.5100 F.S.
Response 1?2sec?3sec
(within±2%F.S)(within±2%F.S.)
Accuracy±1.0%F.S.25±1.5%F.S.25
Temperature coefficient±0.1F.S./1535±0.2F.S./1535
Repeatability±0.5%F.S.
Operating differential
pressure
1
F.S.5SLM 50 300kPa GF.S.50SLM15 300kPa G
F.S.5SLM 100 300kPa GF.S.?50SLM?200?300kPa?G?
Option Medium differential pressure (MP), low differential pressure (LP) specifications are available
Allowable operating pressure 1500kPa G or less
Proof pressure980kPa G
Leak rate1×108Pa/sexcluding permeation of He
Allowable operating temperature545
Allowable operating humidity10 90 RH No condensation allowed
Materials of the gas contact partBody: SUS316
Valve seat :FKM option: CR or NBR
Sealing :FKM option: CR or NBR
Electrical connectionDsub half-pitch 20pin DX10M20S(HRS)
Flow rate setting signal05VDCinput impedance 100k or more)
1or 420mA selectable
Flow rate output signal05VDC external load resistance 250k or more or
4?20mA external load resistance 300 or less selectable
CommunicationRS232C3 wire Or RS4853 wire selectable
Other functionsInstantaneous flow rate integrated flow rate display 7-segment 4-digit LED
display accuracy ±1%integrated display 00009999MAX
Alarm output NPN open collector output 2 points
Max DC35V50mA, pattern setting 5 pattern function
Required power supply24VDC±10% ripple 5% current consumption 300mA max
JointStandard 1/4F9009/16-18UNFStandard 3/8F900 9/16 18UNF
Main unit boreOption 1/8F900, 1/4UJR,Option 1/2F900, 3/8UJR,
?Rc1/4 etc.Rc3/8 etc.
WeightApprox. 1200gApprox. 1500g

(*1)Only for the mass flow controller.