Alicat MCS-Series Anti-Corrosive Gas Mass Flow Controllers

Alicat MCS-Series Anti-Corrosive Gas Mass Flow Controllers rapidly reach setpoints and maintain stable control of mass flow, volumetric flow, or pressure for 128+ gases, including 32 corrosives.

Features of Alicat MCS-Series Anti-Corrosive Gas Mass Flow Controllers

  • Aggressive gas compatibility. Wetted materials compatible with 32 pure corrosive gases and most refrigerant gases.
  • Versatile & accurate. Real-time control with NIST-traceable accuracy to ±(0.4% of reading + 0.2% of full scale) for 128+ gases and 20 custom-defined mixtures.
  • Fast control, multivariate data. Millisecond-response times immediately upon powering up, to control either mass flow, volumetric flow, or pressure and display
  • instantaneous flow, pressure, and temperature readings, in your choice of engineering units.
  • Easy integration. Custom process ports and electrical connectors, and your choice of EtherNet/IP, EtherCAT, DeviceNet, Profibus, PROFINET, Modbus (RTU and TCP/IP), or serial communications for easy integration into your industrial process.

Specifications of Alicat MCS-Series Anti-Corrosive Gas Mass Flow Controllers

  • Available full-scale ranges: 0.5 SCCM to 5,000 SLPM for all gases (10,000 SLPM for H2)
  • Steady-state control range: 1% – 100% of full scale
  • Standard accuracy calibration, NIST-traceable: ±(0.8% of reading + 0.2% of full scale)
  • Optional high-accuracy calibration, NIST-traceable: ±(0.4% of reading + 0.2% of full scale)
  • Repeatability: ±0.2% of full scale
  • Typical control response time: as fast as 30 ms
  • Warm-up time: <1 s

Operating Conditions

  • Compatible gases: up to 130 pre-loaded selectable gas calibrations; additional mixtures definable with COMPOSER™
  • Operating temperature: –10°C to +60°C
  • Full-scale operating pressure: 160 PSIA
  • Max differential pressure: 75 PSID
  • Proof pressure: 175 PSIA
  • User selectable reference conditions (defaults): 25°C & 1 ATM (STP) and 0°C & 1 ATM (NTP)

Included with each device

  • Multi-variable process measurements: Simultaneously measure mass flow, volumetric flow, gas pressure, and gas temperature.
  • User changeable units of measurement: Change the unit of measurement for pressure, flow, or temperature from our extensive list of engineering units.
  • Gas selectability: Select your process gas from any of the 98+ preloaded NIST-traceable gas calibrations in our gas list.
  • Gas Select COMPOSER™: Define 20 custom gas mixtures of up to 5 constituent gases using our Gas Select Composer™ firmware.
  • Monochrome backlit LCD with integrated touchpad: Easily read measurements using the backlit display and control configurations with the user-friendly integrated touchpad.
  • User-adjustable setpoint ramping & limits: Configurable ramp rates allow the process variable to be controlled as fast or as slow as desired, and configurable setpoint limits protect processes from inadvertent setpoint changes.
  • User-adjustable standard temperature and pressure (STP) conditions: Quickly change reference conditions without need for recalibration.
    Valve close override and valve open override: Utilize alarms to quickly vent your system or stop all flow as needed.