O2 Monitors for Glove Boxes and Nitrogen Generators designed specifically for measuring oxygen in glove boxes from air (21% O2) down to low levels (0-10 ppm O2). This oxygen monitor range has easy installation options that ensure the best fit with the glove box operators needs.
The analyzer or remote sensors can be connected directly on the glove box using a KF-40 flange. In the remote version the oxygen sensor can be placed in the glove box using our sample/calibration module.
O2 Monitors Highlights
- Measurement ranges from 0-10ppm up to 0-25% O2
- Accuracy of better than 2% of selected range
- 24 months sensor life span (in normal use)
- 4-20 mA output
- 18-24V loop powered
- XLT sensor options for CO2 backgrounds
GPR-1500 GB
For trace oxygen measurements as low as 0.1 ppm O2 in various background gases. The sensor is housed in a stainless steel case that can be screwed into place.
Options :
- Calibration module : Special stand for sampling or calibration (SS-3170)
- Available ranges : 0-10, 0-100, 0-1,000 ppm & 0-1% O2, 0-25% for calibration only
GPR-2500 GB
For oxygen measurements from 21% down to 100 ppm O2 in a range of gas backgrounds.
Options :
- Calibration module : Special stand for sampling or calibration (B-3170)
- Available ranges : 0-1, 0-5, 0-10% and 0-25%
O2 Monitors Features
- Measurement ranges from 0-10ppm up to 0-25% O2
- Accuracy of better than 2% of range
- 24 months sensor life span (in normal use)
- 4-20 mA output
- 18-24V loop powered
- XLT sensor options for CO2 backgrounds
Nitrogen Generators Applications
- Monitoring oxygen in glove boxes
- Oxygen depletion in confined spaces (GPR-2500 only)
- Measuring oxygen in nitrogen generators
Nitrogen Generators Specifications
GPR-1500 GB | GPR-2500 GB | |
Measurement range | 0-10, 0-100, 0-1000 ppm, 0-1%, | 0-1%, 0-5%, 0-10%, 0-25% |
0-25% (calibration only) | ||
Accuracy | < 2% of selected range at constant conditions | |
Response time | T90 < 10 seconds | |
Recovery time | 60 seconds in air to < 10 ppm in < 1 hour on N2 purge | Not applicable |
Sensitivity (LDL) | 0.05 ppb | 0.005 % |
Linearity | < 1% of scale | |
Sensor model | GPR-12-333-M | GPR-11-32-4 |
XLT-12-333-M for gases containing > 0.5% CO2 | XLT-11-24-4 for gases containing > 0.5% CO2 | |
Sensor life at 25ºC (77°F) and 1 atm | 24 months in < 1000 ppm O26 months in air | GPR-11-32-4 32 months in air |
XLT-11-24-4 24 months in air | ||
Calibration interval | 30 days | |
Inlet pressure | Nominally atmospheric | |
Flow through system: 0.34-2 barg (5-30 psig) with atmospheric vent | ||
Flow rate | Ambient monitoring or Flow through system 0.5-1 Nl/min (1-2 SCFH) | |
Gas connections | KF 40 flange or 1/8” compression tube fittings (with flow through housing) | |
Wetted parts | GPR-1500 GB Stainless steel GPR-2500 GB Delrin | |
Display | Graphical LCD 7 x 3.5cm (2.75 x 1.375″); resolution 0.001 | |
Enclosure | Painted aluminum, 7.6 x 10.1 x 5.1cm (3 x4 x2″) | |
Weight | 900g (2lbs) | |
Compensation | Temperature | |
Signal output | 4-20mA (loop current) | |
Alarms | None | |
Operating temperature | GPR sensor: 5°C to 45°C (41°F to 113°F) | |
XLT sensor: -10°C to 45°C (14°F to 113°F) | ||
Power | 18-24 V DC two wire loop | |
Area classification | General purpose |
Download Datasheet : Analytical Industries Oxygen Monitors for Glove Boxes and Nitrogen Generators
Reference: www.processsensing.com