Microsensor Corp MDM4901FL Differential Pressure Transmitter

Microsensor Corp MDM4901FL Piezoresistive Differential Pressure Transmitter Pressure Transmitter is dedicated to oxygen pressure measurement, filled with Fluorinated® oil, which has high precision, high stability, and high safety level. Microsensor Corp Pressure Transmitter is widely used in the medical and pharmaceutical industries, specializing in gas pressure measurement in medical oxygen production equipment and oxygen storage tanks.

Microsensor Corp MDM4901FL Features

  • Full stainless steel construction, compact size, easy installation
  • Laser welding, fully-sealed construction; protection IP65
  • Using piezoresistive differential pressure sensor, 316L isolated diaphragm
  • Temperature compensation and aging, stable performance
  • Zero and span adjustable outside
  • Ex-proof version MDM4901FL conforms to GB3836.4 Exia Ⅱ CT6 Ga standard; ex-proof certificate is approved
  • CE and ROHS certificates

Electric Performance

  • Power supply: 15V~28V DC
  • Output signal: 2-wire 4mA~20mA DC; 3-wire 0V~5V DC
  • Electrical connection: plug connection or Φ7.2mm 7-pin cable
  • Response time(10%~90%): ≤1ms
  • Insulation resistor: 100MΩ,500V DC

Environment Condition

  • Shock effect: ≤1% at 3gRMS, 30Hz~2000Hz
  • Impact: ≤1% at 100g, 10ms
  • Lifetime: 1×108 pressure cycles
  • Media: liquid or gas which is compatible with construction material

Construction Performance

  • Housing: stainless steel 304
  • Diaphragm: stainless steel 316L
  • O-ring: EPDM
  • Filled liquid: Fluorinated® oil
  • Pressure port: G1/4 female

Microsensor Corp MDM4901FL Specifications

Range code 0A 02 03 07 08 09 10 12 13
Unit bar
Measure range 0~0.35 0~0.7 0~1 0~2 0~3.5 0~7 0~10 0~20 0~35
+overpressure 0.7 1.5 2 4 7 14 20 40 70
-overpressure 0.35 0.7 1 2 3.5 7 10 10 10
Max.static pressure ≤200bar
Item Min. Typ. Max. Unit
 Accuracy 0bar ~ 1bar 0.5

 %FS

2bar ~ 35bar 0.25 0.5
 Zero Thermal error 0bar ~ 1bar 0.75 1.25

 ±%FS, @25℃

2bar ~ 35bar 0.5 0.75
 FS Thermal error 0bar ~ 1bar 0.75 1.25
2bar ~ 35bar 0.5 0.75
 Stability ≤2bar 0.5

 %FS/year

≤35bar 0.2
Static pressure effect 0.05 ±%FS, each 1bar
Compensation temp. 0~50

 

Operation temp. -30~80 ; -10~70(Cable)
Storage temp. -40~120; -20~85(Cable)

Notes

  1. We suggest to install tri-valve between the measured point and transmitter to protect the media adding on transmitter’s positive and negative cavities slowly;
  2. We suggest to make two pressure ports horizontally to reduce installation direction effect;
  3. Please pay attention that the static pressure should be less than 200bar, transmitter positive and negative cavity should be in the rating pressure range;
  4. Please note ex-proof, M6 or M7 options in the order if the user needs ;
  5. Digital indicator information, please refer to MPM4901FL datasheet;
  6. If the user has special requirement, please feel free to contact our

Download Datasheet: MDM4901FL, Microsensorcorp-Piezoresistive Differential Pressure Transmitter

Ref: Microsesorcorp.com