Aquaread GPS Leveline Meter

Overview

Aquaread GPS Leveline Meter is designed to be used with all Leveline Water Level and Temperature loggers. The meter has many uses including LeveLine setup, data download, and the novel GPS embedding feature which tags the LeveLines deployment location into the dataset

Aquaread GPS Leveline Meter Features

Easy to use interface

  • Full LeveLine setup directly from the GPS LeveLine Meter
  • Initiate Scheduled logging meaning there is no need to set a planned start date
  • Embed the LeveLine with its GPS deployment coordinates
  • Download data from multiple LeveLines deployed in the field
  • Perform live spot testing
  • Check LeveLines battery and memory levels prior to deployment
  • Manually input the level for salinity compensation
  • Built in air pressure sensor for atmospheric pressure measurement
  • Ergonomic curved design sits nice in the hand, with rubber gasket for added grip in wet condition
  • Rugged, metal AquaConn connectors provide extreme durability where its needed most
  • LeveLine PC application and USB cable provided with every PC-KIT purchased for use with the LeveLine

Applications

  • Groundwater level monitoring, pump tests, slug tests etc.
  • Stream, lake and reservoir water level measurement.
  • Wetland and flood water monitoring.
  • Coastal monitoring.
  • Tank level measurement.
  • Long term continuous monitoring in boreholes, surface water and seawater applications.

MAKE YOUR LIFE EASIER IN THE FIELD

Some of the GPS LeveLine Meters features can really assist your logger deployment and your logger collection when the scheduled logging is complete. The meter allows you to make adjustments to your LeveLine in the field. For example you can increase the logging duration or increase the logging frequency. In fact you can perform all of the set up actions that you can on the LeveLink PC application.

When you are on site and ready to deploy your LeveLine you can use the GPS LeveLine Meter to check the LeveLine’s battery and memory levels, you can also zero the depth measurement against the current reading of air pressure measured by the meter.

EMBED THE LEVELINE WITH ITS GPS DEPLOYMENT COORDINATES

The most novel feature of the GPS LeveLine meter is the ability to embed the GPS deployment coordinates into the LeveLine. This means that when you collect your loggers and take them back to the office, each logger knows exactly where it was deployed. The information is contained within the logger and can be displayed in Google Earth.

Aquaread GPS Leveline Meter Specifications

LEVELINE (Abs & Gauge)LEVELINE – BARO
GeneralTemperature ranges (non freezing)Operational: -20-80° C (-4-176° F)Operational: -20-80° C (-4-176° F)
Storage: -40-80° C (-40-176° F)Storage: -40-80° C (-40-176° F)
Compensated: -20-80° C (-4-176° F)Compensated: -20-80° C (-4-176° F)
Diameter22mm (0.866 in)22mm (0.866 in)
Length186mm (7.32 in)186mm (7.32 in)
Weight150g (5.3oz)160g (5.6oz)
MaterialsTitanium body, Delrin nose coneTitanium body, Delrin nose cone
Output optionsModbus/RS485, SDI-12, Aquaread proprietaryModbus/RS485, SDI-12, Aquaread proprietary
Battery type & life3.6V lithium; up to 10 years (see note 1)3.6V lithium; up to 10 years (see note 1)
External power6 – 24 VDC6 – 24 VDC
MemorySize8.0 MB2.0 MB
Data Records500150
Log typesLinear, Event & User-Selectable Schedule with Future Start, Future Stop, Deploy Start and Real Time ViewLinear, Event & User-Selectable Schedule with Future Start, Future Stop, Deploy Start and Real Time View
Fastest logging rate & Modbus rate10 per second1 per minute (logging) 5 per second (Modbus)
Fastest SDI-121 per second1 per second
output rate
Accurate to 1 second/24-hr periodAccurate to 1 second/24-hr period (± 6 minutes/year)
Real-time clock(± 6 minutes/year)
Pressure SensorType / MaterialPiezoresistive; ceramicPiezoresistive; ceramic
10.0m (32.8 ft)         20.0m (65.6 ft)
Range (Absolute)50.0m (164 ft),         100m (326 ft)0 to 16.7 psi; 0 to 1.15 bar
Range (Gauge)10.0m (32.8 ft)         20.0m (65.6 ft)N/A
50.0m (164 ft),         100m (326 ft)
Maximum pressureMax 2x range, Burst 2.5x rangeMax 2x range, Burst 2.5x range
Accuracy @ 15° C±0.05% FS±0.1% FS
(see note 2)
Accuracy (FS)±0.1% FS±0.2% FS
( see note 3)
Resolution0.002% FS or 1mm whichever is greater0.1mb
Units of measurePressure: mbar (psi, kPa, bar, mmHg, inHg, cmH2O, inH2O, Level: in, ft, mm, cm and m available in LeveLink)Pressure: mbar (psi, kPa, bar, mbar, mmHg, inHg, cmH2O and inH2O available in LeveLink)
Temperature SensorAccuracy Resolution±0.1° C±0.1° C
0.01° C0.01° C
Output UnitsCelsius (fahrenheit available in LeveLink)Celsius (fahrenheit available in LeveLink)
endent on logging rate. 2) Across factory-calibrated pressure range at a constant temperature. 3) Across factory-calibrated pressure and temperatu