Description
BAMOPHOX 319 ML-DB Bamo Resistivity Meter for process control and survey of demineralized water through two bottles of resins. Input 1: controls the first bottle, as a blind instrument (without display).
As soon as the fluid resistivity is less than the preset value relay 1 switches a signal alarm to inform the operator. Input 2: controls the second bottle with its own alarm parameters. The measurement of resistivity is permanently displayed. 3 relay outputs are to be used either for input 1 or input 2; adjustments are done by the operator. The 4-20 mA outputs allow the report of process operations, in order to have a true image of water quality.
BAMOPHOX 319 ML-DB Bamo Resistivity Meter Features
- 2 independent sensors
- Input 1: range 0…20 M?
- Input 2: scalable range 0…20 M? / 0…2 M?
- Dedicated probes and sensors
- With Factor probe C = 0.01, 0.1, 1 and 10
- 2 adjustable analog output 0/4-20 mA
- 3 relay outputs (NO contact)
- Wall mounting IP65
Specification
Displayed parameters: | Measurement values – Configuration Menu |
Display: | Backlighted – 1 line of 16 alphanumerical characters; 9,2 mm high |
Indication: | LED alarms status |
Configuration: | 8 push buttons keyboard on front face – Keyword protected |
Scales: | 200 ? to 200 M? |
Accuracy: | ± 0,3% |
Probe input: | Coaxial connector BNC type |
Input 1: 20 M? | |
Input 2: Adjustable | |
Temperature compensation: | On range 20 M?, automatic with an input for a 3 wires Pt 100 ? /0° |
Manually from 0 to 100°C | |
CAUTION: with 0.1 sensor Factor there is no automatic compensation | |
Relay outputs: | 3 closing contacts (Silver alloy), voltage free |
Thresholds: | 3 programmable independent thresholds |
– with adjustable hysteresis 0…100% | |
– and adjustable timer from 0 to 9999 sec | |
Output relay (S4): | Not available |
Contact: | Initial resistance 100 m? as a maximum (voltage drop 6 V DC 1 A) |
Rated at 831 V AC / 3 A / 277 V AC; 90 W / 3 A / 30 V DC | |
Switching capacity (minimum) 100 mA, 5 V DC (depending of switching conditions, accuracy) | |
Mechanical life time (minimum) 5 x106 operations (180 commutation/min) | |
Electrical life time (minimum) 2 x105 (20 comm./min) [3 A, 125 V AC], [3 A, 30 V DC] and 105 (evaluated charge) for 3 A, 125 V AC | |
Calibration sequence: | Relay outputs are inhibited, and analogical outputs stand on the last values |
Measurement output: | 0/4-20 mA (maxi 600 ?) proportional to measurement, galvanic insulated |
Program Testing: | Simulation through the menu on measurement and relay outputs |
Main power supply: | 230 V AC / 50-60 Hz [other on request] – Consumption 10 VA |
Models: | Wall mounting, IP65, cable glands, connections on screw terminal |