Description of Gastec Gastec WP-135 Waste Water Test Kit
Enables concentrations of organic chlorinated solvents (e.g. tetrachloroethylene, trichloroethylene) present in wastewater to be easily measured. *We also have the Wastewater test kit WPT-132 / WPT-133 / WPT-135 which includes the Gas sampling pump GV-100 instead of the Plastic syringe pump.

Detector Tube
The substance to be measured | Tube No. & Name | Measuring range | |
Trichloroethylene | 132M | Trichloroethylene | 0.14 – 7.0 mg/l |
132LL | Trichloroethylene | 0.007 – 0.256 mg/l | |
132L | Trichloroethylene | 0.056 – 1.95 mg/l | |
132LC | Trichloroethylene | 0.1 – 0.4 mg/l | |
Tetrachloroethylene | 133M | Tetrachloroethylene | 0.13 – 6.3 mg/l |
133LL | Tetrachloroethylene | 0.005 – 0.174 mg/l | |
133L | Tetrachloroethylene | 0.052 – 1.93 mg/l | |
133LC | Tetrachloroethylene | 0.05 – 0.3 mg/l | |
1,1,1-Trichloroethane | 135L | 1,1,1-Trichloroethane | 0.4 – 14 mg/l |
How to Use Gastec WP-135 Waste Water Test Kit
- Measure 200mL of wastewater with the beaker, replace it in the glass jar, seal the opening with a stopper, and shake it for one minute.
- Both ends of the detector tube are broken off, and it is inserted in the syringe or sampling pump (GV-100) for gas sampling.
- The stopper of the glass jar is removed, and measurement is done with the point of the detector tube just above the surface of the liquid.
- After the measuring time is over, the scale on the detector tube is read.
- The temperature of the wastewater is measured.
- By using the conversion data of the user manual, calculate the concentration in the wastewater.
For more details information and price, please click this link: Gastec WP-135 Waste Water Test Kit