ABB Multivariable Transmitter type 266CRH and 266CSH

ABB Multivariable Transmitter with remote diaphragm seals calculates the mass flow for gases, vapors, and liquids with dynamic flow compensation directly from the three process variables of differential pressure, absolute pressure, and temperature.

The multivariable transmitters in the 266CRH are used for differential pressure flow measurements. Apart from pressure- and/or temperature-dependent changes to the density of the fluid, the parameters such as the discharge coefficient, thermal expansion of pipelines and primary device, and the Reynolds number are also corrected. It can also be used for DP level measurement if density compensation is necessary due to process temperature changes. This could make any level measurement much more reliable.

In addition to high precision, the multivariable 266 also offers the advantage that only a single device needs to be used for the measuring point instead of several transmitters and a flow calculator.
266CRH
266CRH

Specifications | Technical Data

  • Base accuracy: ± 0.075 %
  • Reduced process penetrations save money and reduce the risk of leakage
  • High static working pressure option (41 MPa, 6000 psi) for power plants, pipelines or oil and gas production
  • Local operation with LCD with TTG (Through The Glass) technology without the need of opening the housing
  • Integrated totalizers and in addition a binary output as pulse or frequency output for activating external totalizers
  • Exchangeable electronics with automatic configuration capability
  • Flexible process connections and high selection of corrosion-resistant material due to all welded diaphragm seals

ABB Multivariable Transmitter 266CSH

The multivariable transmitters in the 266CSH calculate the mass flow for gases, vapors and liquids with dynamic flow compensation directly from the three process variables of differential pressure, absolute pressure, and temperature.

ABB Multivariable Transmitter in the 266CSH are used for differential pressure flow measurements. Apart from pressure- and/or temperature-dependent changes to the density of the fluid, the parameters such as the discharge coefficient, thermal expansion of pipelines and primary device, and the Reynolds number are also corrected. It can also be used for DP level measurement if density compensation is necessary due to process temperature changes. This could make any level measurement much more reliable.

In addition to high precision, the multivariable 266 also offers the advantage that only a single device needs to be used for the measuring point instead of several transmitters and a flow calculator.

ABB Multivariable TransmitterSpecifications | Technical Data

  • Base accuracy: ± 0.075 %
  • Reduced process penetrations save money and reduce the risk of leakage
  • High static working pressure option (41 MPa, 6000 psi) for power plants, pipelines or oil and gas production
  • True draft range (down to 0.05 kPa, 0.2i nH2O) transmitter
  • Local operation with LCD with TTG (Through The Glass) technology without the need of opening the housing
  • Detection of plugged impulse lines
  • Integrated totalizers and in addition a binary output as pulse or frequency output for activating external totalizers
  • Exchangeable electronics with automatic configuration capability

Download Datasheet: 266CRH