BAMOSOFT Control Bamo Water Hardness Monitor, monitors safely the quality of softened water. It emits a light alarm as soon as the hardness of outlet water is near or above the threshold. (See the signaling table)
Example : Raw water (tap water): hardness of 300 ppm CaCO₃ Threshold is set at 0.25 %: for alarming when hardness of softened water is above 0.25 % of raw water hardness. It means that an alarm occurs when softened water hardness is over 0.75 ppm of CaCO₃.
BAMOSOFT Control also delivers a 4-20 mA signal which represents the percentage of residual hardness in the softened water, in comparison to raw water hardness.
Example : Raw water hardness (tap water) is 300 ppm CaCO₃. Output signal 4-20 mA is set on the scale 0 … 2 %. The value 20 mA will represent 2% of raw water hardness, here: 6 ppm CaCO₃.
BAMOSOFT Control Operation
BAMOSOFT Control comprises two solenoid valves that successively pass raw water and then softened water on a specific Ca⁺⁺ /Mg⁺⁺ electrode. With its hardware, the analyzer contemplates the ionic activities, compares data and proceed to check the system, less or more frequently. The controller delivers information: normal operation, alarm, faulty solenoid valve, brine leakage, etc.
Water samples are rejected to drain (no pressure). After start up, once measurements are stabilized, BAMOSOFT triggers a sampling routine each 15 minutes. It is possible to trigger sampling by an external signal (counter with pulse, external contact).
Signaling table of LEDs
Bamo Water Hardness Monitor Depending on their meanings, they turn on “continuous or flashing. In the table below, S stands for “Threshold set”.
Features
- Fit any water softener system
- On-line control and monitoring
- Alarm set up from 0.25% to 10% of raw water hardness
- Self diagnostic of complete system
- Signal outputs for remote information
Application
Control and monitoring of softened water quality, by comparison of waters from softener inlet and outlet.
Another Article: |
BAMOSOFT Control Specification
Water temperature | Water: 5 … 30 °C; Ambient: 5 … 50 °C |
%R.H. | 20 … 80 % RH |
Power Unit | 100 … 240 V – 50/60 Hz / 0.2 A |
Output delivered voltage | 15 V DC; approx. 6 W |
Relay output | 2 Change-over contacts; max. 24 V DC / 0.5 A |
Relay 1: Threshold over-passed or brine detected | |
Relay 2: Electrode fault or Solenoïd valve fault | |
Fittings for samples | By-pass with fittings for PA tube |
PA tube: O.D. 4 mm, i.d. 2 mm ; In-line filter < 0.1 mm is highly recommended | |
Samples pressure limits | Min. 1 bar; Max. 7 bar |
Outlet to drain | PA Tube, O.D. 6 mm; No pressure |
Raw water quality | Drinkable quality, free of grease, oil and brine |
Restrictions | Concentration of iron and manganese less than 0.05 mg/l (More information: Contact us) |
Hardness range | 10 to 1000 ppm CaCO₃ |
Trigger point | 8 Thresholds in % of raw water hardness: 0.25 ; 0.5 ; 1 ; 2 ; 4 ; 6 ; 8 or 10 % |
Analogue output signal | 4-20 mA (max. 250 Ω) – Scales 0 … 2% or 0 … 10% of raw water hardness |
Electrode check-up | Self-diagnostic with alarm for low effectiveness |
External sampling signal | Pulse limit > 0.5 s |
Dimensions | Wall mounting; Case: 95 x 145 x 35 mm (Overall 120 x 220 x 120 mm) |
Download Data Sheet: Water Hardness Monitor BAMOSOFT Control
Ref : Bamo
You must be logged in to post a comment.